Comparative Analysis Between Lumen Method Calculation and Dialux Simulation for Electrical Lighting Design.
DOI:
https://doi.org/10.37933/nipes.e/4.3.2022.7Abstract
This paper presents a comparative analysis between Lumen method
and Dialux simulation for lighting design for an office block. The
analysis is to determine which method gives the number of
luminaires which produces the recommended photometric
parameters by Illumination Engineering Society of North America
(IESNA) standards for an office. The number of luminaires required
for the office block was computed with Lumen method and Dialux
Evo 9.2 using Spectral MIREFA luminaires specifications. Results
was simulated on Dialux Evo 9.2 to determine the photometric
parameters; illumination, uniformity of illumination and Unified
Glare Rating (UGR). The Lumen method had one luminaire more
than the Dialux simulation. Results from the photometric analysis
showed that for a target value of 300lx, the Lumen Method achieved
a better perpendicular illuminance of 367lx, 347lx and 314lx against
213, 248 and 241 from Dialux. Also, the Lumen method achieved
better uniformity of illumination, having values of 0.034, 0.020 and
0.030 against 0.074, 0.002 and 0.024 gotten from Dialux design for
the small, medium and large offices respectively, though both results
were not up to the standard value recommended by IESNA as 0.4.
The UGR results for Lumen method had points with stronger glare
exceeding the standard of 19 set by IESNA, while Dialux had lower
glare, making it the better design in terms of glare rating. Overall
comparism shows that the Lumen method is better suited for lighting
design calculations using Spectral MIREFA and Philips RC540B
lighting fittings, while Dialux is suitable for lighting simulation and
presentation.